The digital dust particle counter features fast response and stable performance, supporting real-time monitoring and routine inspections. By combining particle size detection with environmental parameter measurement, it provides a practical solution for professional air quality monitoring and maintenance tasks.

High-Precision Multi-Size Detection

WiFi Connectivity for Remote Monitoring
The particle counter is a precision instrument used to measure the number and size distribution of suspended particles in the air. The application of particle counters not only improves production efficiency and product quality, but also helps to protect the health and safety of workers, and also plays a positive role in environmental protection. It has a wide range of applications in many fields.
| Model | SISCO-PC-HZ-012 |
| Particle Size Channels | 0.3 µm, 0.5 µm, 1.0 µm, 2.5 µm, 5.0 µm, 10 µm |
| Measurement Range | Particle: ≤ 2,000,000,000 PCS/m³; ≤ 2,000,000 PCS/L |
| Differential Pressure: ±36 kPa | |
| PM2.5 / PM10: 0–1000 µg/m³ | |
| CO2: 0–10,000 ppm | |
| Temperature: −20 to 60 °C | |
| Humidity: 0–100 %RH | |
| Accuracy | Differential Pressure: ±0.3% FSO |
| CO: ±(50 ppm + 3% of reading) | |
| PM2.5 / PM10: 0–100 µg/m³±10 µg/m³, 100–1000 µg/m³±10% | |
| Temperature: ±2°C | |
| Humidity: ±3 %RH | |
| Sampling Flow Rate | 0.1 ft³ (2.83 L/min) |
| Cleanliness Standard | ISO 14644-1: ISO Class 5 – ISO Class 9 |
| Response Time | 1 s |
| Stabilization Time | 60 s |
| Internal Memory | 1 GB |
| WiFi APP | Tuya Smart |
| Power Supply | 3.7 V / 10,000 mAh polymer lithium battery |
| Display | 4.0-inch TFT capacitive touchscreen |
| Continuous Operating Time | ≥ 8 h |
| Charging Time | 2–3 h |
| Data Storage | 10,000 records (customizable) |
| Net Weight | 1000 g |
| Product Dimensions | 231 × 88 × 56 mm |
| Package Dimensions | 275 × 225 × 97 mm |
| Gross Weight | 2000 g |
Q1: How to use the handheld particle counter?
A1: The particle counter is simple to use, first determine the number of test points according to the area of the clean room and distribute the test points evenly in the clean room, then set the measurement time, measurement unit, flow rate, and other parameters of the particle counter, then put the particle counter on the plane of 1~1.2 meters height from the ground in the middle of the sampling point, and finally start to test the sampling and record the data one by one.
Q2: How to change the printing paper of the particle counter?
A2: Generally, the printing part of the particle counter has a cover, some covers can be lifted directly to cover, lift it to replace the printing paper, and some need to use a screwdriver to unscrew the screw, the specific instructions are also introduced, open the cover to follow the thermal printing paper can be replaced.
Q3: How to read the dust particle counter?
A3: The results of the dust particle counter, generally the left column is the particle diameter, and the right column is the number of particles per cubic meter.
Tips: What do the readings on the portable air particle counter mean?
The readings show how many particles are in the air for each particle size. Higher numbers mean there is more dust or pollution in the air, while lower numbers indicate cleaner air. Each size channel tells you how many particles are equal to or larger than that size.
Thank you for buying industrial test and measurement equipment on SISCO.com, all products sold by SISCO and the partner cover a 12 months warranty, effective from the date of receiving the products.
SISCO is responsible for providing free spare parts, and free technical support to assist the customer to repair the defective products until the problem is solved.